Back to Products
S230 / S231 Dew Point Sensor (-100 … +20 °C Td)

S230 / S231 Dew Point Sensor (-100 … +20 °C Td)

The SUTO S230/S231 Dew Point Sensors provide reliable dew point monitoring for industrial and hazardous applications. SUTO’s unique dual sensor technology optimizes sensor sensitivity and accuracy by automatically selecting the ideal sensor type for the situation.

  • Useful: measure down to -100 °C Td
  • Precise: advanced QCM sensor technology
  • Integrated: easy setup with SUTO data loggers

Description

S230 / S231 Dew Point Sensor (-100 … +20 °C Td)

Description

The SUTO S230 / S231 dew point sensors provide reliable, long term stable dew point monitoring for hazardous industrial applications. SUTO’s unique dual sensor technology automatically selects between two sensors included in both the S230 and S231, providing higher levels of sensitivity and accuracy. The S230 / S231 comes ready-to-use with your choice of 4-20mA or Modbus/RTU (RS485) outputs. If required, parameters can quickly and easily be configured through the SUTO service software.

Benefits

Benefit Icon

Never Worry About Accuracy

Automatic dual sensor technology ensures accurate readings

Benefit Icon

Sensors You Can Count On

Rugged IP65 casing protects from even the harshest conditions

Benefit Icon

Constant Monitoring

Catch minor issues before they cause real problems

Features

Features

  • ATEX, IECE X and GB EX Approval
  • Low Dew Point: Measures down to -100 °C Td
  • Dual Sensor System: High precision over the whole range
  • Precise Measurement: Unique QCM sensor technology

Specs

Specs

  • Measurement range (model depending):
    • Dew point
      • -50 … +20 °C Td
    • Temperature
      • -30 … +70 °C
  • Dew point sensor:   QCM & Polymer
  • Temperature sensor:  NTC
  • Pressure sensor:   N/A
  • Accuracy:
    • Dew point:   ±2 °C Td
    • Temperature:   0.3 °C
  • Operating Pressure (model depending):   -0.1 … 35 MPa (S231)
  • Operating Temperature (Medium):  0 … +50 °C
  • Measured gasses (Medium):  Non-corrosive gasses
  • Response Time t90 (@ 4 l/min):
    • -20 °C Td -> -60 °C Td = < 240 sec
    • -60 °C Td -> -20 °C Td = < 30 sec
  • Ambient Temperature:   0 … +50 °C
  • Ambient Humidity:   0 … 100% rH
  • Supply Voltage:   12 … 30 VDC
  • Current consumption:   40 mA @ 24 VDC
  • Output signals:  4 … 20 mA (isolated) Modbus/RTU
  • Electrical connection:   Screw terminals
  • Process connection:   G 1/2“ thread (ISO 228/1) Stainless steal 1.4301 (SUS 304)
  • Casing material:   Aluminum alloy
  • Classification:   IP67
  • EMC:   IEC 61326-1
  • Approval:   Ex db[ib] IIC T4 Gb
  • Sensor protection:   Sinter filter
  • Transport Temperature:   -30 … +70 °C
  • Storage Temperature:   -20 … +50 °C
  • Weight:   728 g

Documents

S230_231-Catalog

Catalog Cover S230_231-Catalog (693 KB)

S230_231-Manual

Catalog Cover S230_231-Manual (1 MB)